Simultaneous determination of ondansetron and tropisetron in human plasma using HPLC with UV detection.
Journal: 2002/July - Biomedical Chromatography
ISSN: 0269-3879
Abstract:
A rapid and sensitive HPLC method for the simultaneous quantitation of ondansetron and tropisetron, two serotonin (5-HT) receptor antagonists frequently used in treatment and prevention of nausea and emesis, is described. The procedure involves liquid-liquid extraction of human plasma with dichloromethane coupled with reversed-phase HPLC and UV detection. The lower limits of quantification (LOQ) were 0.62 ng/mL for ondansetron and 1.25 ng/mL or tropisetron. Intra- and inter-assay coefficients of variation ranged from 1.5 to 7.5% and 5.3 to 13.7%, respectively. The sensitivity and precision were sufficient for determination of plasma concentrations after therapeutic administration of both drugs and the method can be used for the estimation of pharmacokinetic parameters.