Three-dimensional analyses of ultrasonic scaler oscillations.
Journal: 2009/June - Journal of Clinical Periodontology
ISSN: 1600-051X
Abstract:
BACKGROUND
It is stated that the oscillation patterns of dental ultrasonic scalers are dependent upon whether the instrument is of a magnetostrictive or piezoelectric design. These patterns are then linked to differences in root surface debridement in vitro.
METHODS
Piezoelectric (A, P) and magnetostrictive (Slimline, TFI-3) ultrasonic scalers (three of each) were evaluated, loaded (100 g/200 g) and unloaded with a 3D laser vibrometer. Loads were applied to the probe tips via teeth mounted in a load-measuring device.
RESULTS
Elliptical motion was demonstrated for all probes under loaded and unloaded conditions. Loading flattened the elliptical motion along the length of the probe. Unloaded, Slimline tip 1 was significantly different to tips 2 and 3 (p<0.0001). There were no differences between the A-tips (p>0.207). All TFI-3 tips were different to each other (p<0.0001). P-tips 1 and 2 were different to each other (p=0.046). Loaded, Slimline tips were different to each other (p<0.001). There were no differences between the P probes (p>0.867). Generator power increased all Slimline and P tip vibrations (p<0.0001).
CONCLUSIONS
Probe oscillation patterns are independent of ultrasound production mechanism and are dependent upon probe shape and generator power. Loaded probes oscillated with an elliptical pattern.