Use of covered stents in simultaneous management of coarctation of the aorta and patent ductus arteriosus.
Journal: 2018/December - Anatolian journal of cardiology
ISSN: 2149-2271
Abstract:
OBJECTIVE
To report clinical and procedural characteristics of twelve patients who received a covered stent for the treatment of aortic coarctation and concurrent patent ductus arteriosus (PDA).
METHODS
A single center database was retrospectively evaluated to obtain data of patients with combined aortic coarctation and PDA. We selected patients in whom a covered stent was used for the treatment of both pathologies. The stent length was chosen so as to cover the entire length of the lesion from healthy to healthy tissue and also cover the ampulla of PDA.
RESULTS
The median age of the patients was 15 (range, 6.5-35) years. The diameter of the coarctated segment increased from a median of 8.4 (range, 2.6-10.8) mm to 16 (range, 9-24) mm (p<0.005), whereas the pressure gradient decreased from a median of 43 (range, 10-71) mm Hg to 0 (range, 0-8) mm Hg (p<0.005). Fourteen covered stents were used for 12 patients. Following deployment, seven stents were flared with larger and low-pressure balloons because of the gap between the distal end of the stent and the poststenotic dilated segment of the aorta, which caused residual PDA shunts and/or instability of the stent. After the procedure, no residual PDA shunt was present in any patient.
CONCLUSIONS
To the best of our knowledge, this study includes the largest series of patients reported in literature in whom covered CP stents were used for simultaneous percutaneous treatment of coarctation and PDA. The procedure was successful and stable results were obtained during follow-up in all cases.
Relations:
Content
Citations
(1)
Similar articles
Articles by the same authors
Discussion board
Anatolian Journal of Cardiology. Mar/31/2018; 19(4): 232-236
Published online Mar/20/2018

Use of covered stents in simultaneous management of coarctation of the aorta and patent ductus arteriosus

Abstract

Objective:

To report clinical and procedural characteristics of twelve patients who received a covered stent for the treatment of aortic coarctation and concurrent patent ductus arteriosus (PDA).

Methods:

A single center database was retrospectively evaluated to obtain data of patients with combined aortic coarctation and PDA. We selected patients in whom a covered stent was used for the treatment of both pathologies. The stent length was chosen so as to cover the entire length of the lesion from healthy to healthy tissue and also cover the ampulla of PDA.

Results:

The median age of the patients was 15 (range, 6.5-35) years. The diameter of the coarctated segment increased from a median of 8.4 (range, 2.6-10.8) mm to 16 (range, 9-24) mm (p<0.005), whereas the pressure gradient decreased from a median of 43 (range, 10-71) mm Hg to 0 (range, 0-8) mm Hg (p<0.005). Fourteen covered stents were used for 12 patients. Following deployment, seven stents were flared with larger and low-pressure balloons because of the gap between the distal end of the stent and the poststenotic dilated segment of the aorta, which caused residual PDA shunts and/or instability of the stent. After the procedure, no residual PDA shunt was present in any patient.

Conclusion:

To the best of our knowledge, this study includes the largest series of patients reported in literature in whom covered CP stents were used for simultaneous percutaneous treatment of coarctation and PDA. The procedure was successful and stable results were obtained during follow-up in all cases.

Introduction

The diagnosis and management of aortic coarctation is usually established during infancy or childhood. Occasionally, adolescent or adult asymptomatic patients may be diagnosed during investigations for hypertension. Depending on the patient’s age and the coarctation anatomy, treatment options for aortic coarctation include balloon angioplasty, stent implantation, and surgery.

Patent ductus arteriosus (PDA) is accepted as a congenital heart disease when its persistency continues beyond the third month of life in term infants. Clinical manifestations of PDA vary from asymptomatic heart murmur to congestive heart failure, and even Eisenmenger syndrome. The clinical setting of PDA can be managed medically by percutaneous interventions or surgery.

Aortic coarctation is often associated with PDA. The accepted treatment modality for a combination of these two pathologies is surgery during infancy. However, use of interventional procedures in either sequential or simultaneous sessions has been reported in previous studies (1-3). Use of covered stents as a primary treatment modality for relieving coarctation accompanied by PDA has been reported in older children, adolescents, and adults (4,5). In this study, we report 12 patients who received a covered stent for the treatment of aortic coarctation and concurrent PDA. To the best of our knowledge, this report includes the largest series of patients with concurrent aortic coarctation and PDA treated via a transcatheter route.

Methods

Study cohort

A single center database was retrospectively evaluated to obtain data of patients with combined aortic coarctation and PDA since 2007. We selected only those patients in whom a covered stent was used for the treatment of both pathologies for the study. All patients had a native coarctation, except one who had previously undergone balloon angioplasty for aortic coarctation. Because a covered stent implantation procedure has been deemed suitable for patients weighing ≥30 kg, the study group comprised older children, adolescents, and adults.

Catheterization technique

Informed consent was received from the patients or their legal guardians. All patients underwent catheterization under general anesthesia. Heparin (100 IU/kg, max 5000 IU) was intravenously administered immediately after the right femoral artery and vein were accessed. The coarctated segment was retrogradely crossed with a Judkins right coronary catheter (JR4) and a 0.035-inch floppy guide wire (Terumo®). After placement of a standard 0.035-inch exchange guidewire in the ascending aorta through the JR4 catheter, it was exchanged with a pigtail catheter. Hemodynamic assessment and serial angiograms were obtained using the pigtail catheter. The diameter and length of the stenotic area, diameter of the transverse arch, diameter of the aorta at the level of the subclavian artery, descending aorta diameter at the level of the diaphragm, the minimum diameter and the largest diameter (usually at the aortic ampulla) of PDA, and the length of PDA were measured. The anatomy of PDA was determined according to the classification defined by Krichenko et al. (6).

The profile of the balloon used for implantation was chosen after considering the diameter of the isthmus at the level of the left subclavian artery take-off and so that it would not exceed the diameter of the aorta at the level of the diaphragm. The stent length was chosen so that it would cover the entire length of the lesion from healthy to healthy tissue and also cover the ampulla of PDA. The chosen balloon length was longer than the stent length. In all patients, except two in whom a Z Med balloon (NuMED Inc., Hopkinton, NY, USA) was utilized, Balloon-in-Balloon (BIB) catheters (NuMED Inc., Cornwall, Ontario, Canada) were used. After placing an extra stiff guidewire into the ascending aorta or the right subclavian artery, a long sheath of appropriate size for the profile of the desired balloon catheter was advanced. A covered Cheatham-platinum (CCP) (NuMED Inc., Hopkinton, NY, USA) stent was then manually crimped onto the delivery balloon and advanced through the long sheath. A 90% platinum/10% iridium 0.013-inch wire formed the frame of the CCP stent. It was covered by a polytetrafluoroethylene (PTFE) covering, glued at 0° and 180° at each end of the stent. The expanded PTFE covering could be dilated up to 26 mm before tearing. These stents are available in lengths of 16, 22, 28, 34, 39, and 45 mm. After ensuring that the stent layout was appropriate, the balloon was manually inflated, according to the manufacturer’s recommendations, and the stent was deployed. Angiograms were performed during the stent placement through the side arm of the sheath to evaluate the results and the presence of any dissection or rupture. Hemodynamic studies and serial angiograms were repeated after the procedure. A gap remained between the dilated segment of the descending aorta and the distal end of the covered stent, leading to a residual shunt from PDA. Flaring and sealing the distal end of the stent to the aortic wall using larger low-pressure balloons is required. Heparin’s effect was monitored by activated clotting time and neutralized when it was >200 seconds at the end of the procedure. Manual compression to the groin provided adequate hemostasis for all patients. During follow-up, physical examination, blood pressure measurement, electrocardiography, and echocardiography were performed at every clinical visit at 1, 3, 6, and 12 months. All patients underwent a computed tomographic scan after the first year of intervention.

Statistical analysis

SPSS 18.0 was used (SPSS Inc, Chicago, IL, USA) for statistical analysis. Values are expressed as median (minimum-maximum). Wilcoxon signed-rank test was used for comparison of measurements obtained before and after the procedures. A p-value of <0.05 was considered statistically significant.

Results

The median age of the patients was 15 (range, 6.5-35) years. Six of the patients were males. All patients had upper extremity hypertension before the procedure (the median blood pressure measured from the right arm was 165±15/105±12 mm Hg). Five of the patients were still hypertensive after the procedure and currently use dual antihypertensive medication. The type of aortic coarctation was discrete in all patients, except in one who had a long segment coarctation. All patients had an associated PDA. Six of the PDAs were small, two were medium-sized, and four were large. According to Krichenko’s classification system, six of the PDAs were Type E, four were type C, one was type A, and one was type B (Table 1). The diameter of coarctated segment increased significantly from a median of 8.4 (range, 2.6-10.8) mm to 16 (range, 9-24) mm (p<0.005), whereas the pressure gradient across this segment decreased significantly from a median of 43 (range, 10-71) mm Hg to 0 (range, 0-8) mm Hg (p<0.005). The median fluoroscopy time was 19.5 (range, 12-37) min. Prior to intervention, four patients had associated pulmonary hypertension, which was found to be severe in two patients. In one patient, pulmonary hypertension remained severe after the procedure, and this patient currently uses appropriate medication. Fourteen covered stents were used for 12 patients. The stent lengths ranged from 22 to 45 mm. BIB and Z med balloon catheters were used for stent deployment in ten and two patients, respectively. Following deployment, seven stents were flared with larger and low-pressure balloons because of the gap between the distal end of the stent and the poststenotic dilated segment of the aorta, which caused residual PDA shunts and/or instability of the stent (Fig. 1). After completion of the process, no residual PDA shunt was present in any patient. In one patient, while flaring the stent, a non-progressive extravasation occurred. It persisted on the control CT angiograms after the procedure; hence, the patient underwent another catheterization and received two additional covered stents 20 days after the initial procedure. All patients were stable, without the need for additional intervention during the follow-up period.

Table 1
Demographic characteristics and procedural data
AgeWeightPDAAdditionalPHTPreproceduralPostproceduralBalloonsize BalloonStent typeStent length
(years)(kg)typefindingsSystemic HTSystemic HT(mm)type(mm)
Case 12628CVSD, BAV+++16x4BIBCovered CP45
Case 26.519B--+-10x3Z MedCovered CP34
Case 31655EBAV-++12x4BIBCovered CP34
Case 41238CVSD+ASD++-16x4BIBCovered CP34
Case 53573EBAV-+-18x4BIBCovered CP45
Case 62284E--+-20x3.5BIBCovered CP28
Case 71219EVSD (small)-+-13x3Z MedCovered CP28
Case 82560A--++20x3.5BIBCovered CP34
Case 91026CBAV++-14x3BIBCovered CP22
Case 101553CBAV+SAR+++20x4BIBCovered CP39
Case 111342E--++16x4BIBCovered CP34
Case 121457E--+-18x4BIBCovered CP39
BAV - bicuspid aortic valve; PHT - pulmonary hypertension; HT – hypertension; PDA - patent ductus arteriosus; BIB - balloon-in-balloon; CP - cheatham-platinum; SAR - subaortic ridge; VSD - ventricular septal defect
Figure 1
Angiogram of a patient with a mild to moderate coarctation and large PDA (a); a 34-mm covered stent mounted on a 16-mm balloon was advanced across the coarctation segment (b) and then implanted (c); however, there was significant shunt through the duct after the implantation due to the gap between the distal end of the stent and the poststenotic dilated segment of the aorta (c); postdilation of the distal part of the stent with 20-mm and 23-mm low-pressure balloons (d, e); closed the gap between the aortic wall and covered stent. Finally, there was no residual shunt through PDA (f)

Discussion

Patient age and coarctation anatomy together with the size and shape of PDA are very important variables when considering treatment options for patients with aortic coarctation and concurrent PDA. Generally, the preferred treatment modality for a combination of the two pathologies is surgery during infancy, because the incidence of recoarctation after balloon dilation is significantly higher in infants (7).

Geggel et al. (2) suggested that treatment of infants with aortic coarctation and a small PDA was possible through percutaneous interventions. However, whether these procedures should be performed in the same session or in sequential sessions has not been discussed. The disadvantages of a sequential catheterization for the child and their family include cost and the need for a second hospitalization and a second intervention. However, because balloon angioplasty could disrupt the intimal and medial layers of the aorta, manipulations for PDA occlusion after balloon dilation could increase the risk of aortic dissection, mediastinal hematoma, and aneurysm formation at the newly dilated aortic segment, when performed in single sitting. Moreover, when a balloon angioplasty is performed after coil occlusion of PDA, risk of coil embolization increases significantly because the size of the PDA can change after balloon angioplasty of the aorta (5).

Singhi and Sivakumar (8) reported three different transcatheter strategies for three of their cases with coarctation and PDA. One of the patients was a 13-year-old girl with a severe aortic coarctation and a large PDA; she was treated simultaneously via percutaneous implantation of a covered stent. The second patient was a nine-year-old girl with PDA and a mild isthmus hypoplasia. The PDA was occluded with an Amplatzer® duct occluder. When the pressure gradient across the coarctation rose during the seven-year follow-up, the coarctation was stented. The third patient was a 10-year-old who had PDA and a discrete aortic coarctation. After occlusion of the PDA with an Amplatzer® duct occluder, coarctation was relieved by balloon angioplasty (8).

Simultaneous closure of PDA with dilation of the coarctation can be achieved using covered CP stents. Using a single device in a single sitting would reduce the fluoroscopy and procedure times. Sadiq et al. (4) defined the use of a covered CP stent to treat coexistent aortic coarctation and PDA. Kulkarni et al. (5) reported a 20-year-old girl with severe coarctation, a large PDA, and pulmonary hypertension. By implanting a covered stent during a single session, both pathologies were treated; pulmonary artery pressure decreased, and the patient was discharged on the following day after the procedure (5).

While some physicians prefer using covered stents only for curative purposes, such as following a complication like aneurysm formation or a tear, others prefer the prophylactic use of covered stents to avoid late aneurysm formation. Chang et al. (9) reported 25 cases in whom covered stents were used as the primary modality for the treatment of native coarctation. Five of their cases had associated PDAs that were completely occluded when the coarctation was stented. The risk of occlusion of the side branches of the aorta and spinal artery during implantation of covered stents has been a major concern (10). Because the spinal artery arises below the level of the ninth thoracic vertebra, placing a covered stent at the usual site of coarctation seems safe. Among the major side branches of the aorta, sacrificing the left subclavian artery causes no significant morbidity. Another concern has been the long-term fate of the covered stents because they may need redilation before and after full growth of the patient.

Use of BIB catheters allows proper stent positioning and decreases the risk of migration. The inner and outer balloons of BIB catheters can be inflated separately, and the position of the stents can be checked at that time; hence, a more controlled deployment can be achieved. In our study, for the deployment of stents in the desired position, BIB catheters were used in ten patients. A high-pressure Z Med balloon was used because the correctly sized BIB catheters were not available during the procedures for two patients. We chose the size of the delivery balloon by taking into consideration the diameter of the distal transverse arcus and taking care not to exceed the size.

After deployment of the stents, high-pressure balloons were used for redilation of the residual stenosis. Low-pressure balloons, less than the diameter of the aorta at the level of the diaphragm, were utilized for flaring the stents against the dilated vessel wall. Because the space between the implanted stent and the dilated aortic wall could result in a distal endoleak and a persistent shunt through the ductus, especially in patients with marked poststenotic dilatation, adequate sealing of the stent to the aorta by flaring the stent against the vessel wall with larger in size and low-pressure balloons is essential. Additionally, providing an adequate seal and avoiding residual shunts through PDA, flaring of the stents is advised to decrease resistance within tube grafts and precludes proximal infolding of the stent (11).

All covered stents, except two, were implanted, with excellent acute outcomes. In one patient, aged six and a half years, the stent protruded into the transverse arcus causing a pressure gradient of 10 mm Hg, which did not require reintervention. In another patient, aged 34 years, we noticed a residual flow through the PDA caused by a gap between the descending aortic wall and the distal end of the covered stent. During the flaring of the stent, an extravasation that did not compromise the patient hemodynamically had occurred. After 20 days, two additional covered stents were implanted telescopically to cover the ruptured segment in this patient.

The use of covered stents, instead of bare stents, can reduce the potential complications mentioned above, but not the overall risk. Any tear, dissection, or rupture would be covered because of the PTFE covering of the CCP stents. However, particular attention must be paid to make sure that both ends of the stents are placed over healthy aortic tissue. It should be noted that the covering of the CCP stent does not reach the edges of the stent, leaving uncovered parts that may tear or traumatize the vessel wall when the stent is fully expanded.

Diagnostic cardiac catheterization revealed that two of our patients had severe pulmonary hypertension (patients 1 and 4). Vasoreactivity testing was found to be positive and the Qp/Qs ratio was >1.5 in the patient who had a large ventricular septal defect (VSD) (patient 1). As published previously, PDA closure is advised when the shunt is from left to right and pulmonary vascular reactivity is present in patients with severe pulmonary hypertension (12). After the procedure, an immediate decrease in pulmonary artery pressure was observed. In the follow-up, pulmonary artery pressure, estimated by echocardiographic evaluation via VSD jet, showed that the pulmonary hypertension had been completely resolved. In the setting of a PDA with Eisenmenger syndrome, closure of the duct is not advised because it may cause right ventricular failure (13). The other patient with severe pulmonary hypertension was 12 years old and had aortic coarctation, a large PDA, large VSD, and a medium-sized ASD that had resulted in Eisenmenger syndrome (patient 4). Because the accompanying large VSD, which would permit a bi-directional shunt for the relief of pulmonary artery hypertension, was present, the coarctated segment was dilated and the PDA was simultaneously closed with a covered stent. In this way, both the left and right ventricular pressures and thus the systemic pressure were reduced. In the follow-up, however, medications for pulmonary arterial hypertension were continued.

Aneurysm formation and recoarctation are possible complications after surgical and endovascular repair of aortic coarctation. In our study group, after discharge from hospital, regular evaluation using echocardiography did not reveal any need for reintervention for any of the patients. CT angiography, which was performed after the first year of intervention in all patients, showed mild dilation at the poststenotic segment of the descending aorta in four patients. No other pathology, such as recoarctation or aneurysm formation, was detected in any patient. During a median follow-up period of 85 (range, 12-110) months, no adverse event such as aneurysm formation, stent migration, stent fracture, or dissection was found in any patient.

All patients were hypertensive prior to the interventions. Seven patients became normotensive after the procedure, whereas five patients still needed antihypertensive medications. It is known that hypertension can be an issue during the follow-up period of patients with aortic coarctation even after a successful repair and in the absence of restenosis. Canniffe et al. (14) reviewed 26 articles related to residual hypertension after aortic coarctation repair. They reported the median prevalence of hypertension after coarctation repair to be 32.5% (range, 25%-68%) in this review of 26 studies. Our results with residual hypertension were consistent with these findings: Five patients (41.6%) were still hypertensive after the procedures. Although age at the time of repair, present age, method used to measure blood pressure, and description of the hypertension all influence the prevalence, hypertension continues to be a common and potentially serious problem even after successful repair (14).

Study limitations

The retrospective design was the main limitation of the present study. Additionally, the sample size was relatively small due to restriction of the population to older children, adolescents, and adults as a result of the requirement of larger sheaths for covered stent implantation.

Conclusion

To the best of our knowledge, this study includes the largest series of patients reported in literature in whom covered CP stents were used for simultaneous percutaneous treatment of coarctation and PDA. The procedure was successful and stable results were obtained during follow-up in all cases. We believe that if stent implantation is possible, use of a covered stent enables treatment of both conditions in one setting, is safe and simple, and avoids use of an additional device to close the ductus. Besides, the fluoroscopy and procedure times as well as the number of future interventions reduced. We find the use of BIB catheters convenient in this subgroup of patients because these allow proper stent positioning and decrease the risk of migration. We should keep in mind that high-pressure balloons may be needed for further dilation after the deployment to close residual shunts through PDA.

Footnotes

Conflict of interest: None declared.

Peer-review: Externally peer-reviewed.

Authorship contributions: Concept – E.H.Y., M.O.B., M.K., İ.K.Y., A.E., A.Ç.; Design – E.H.Y., M.O.B., M.K., İ.K.Y., A.E., A.Ç; Supervision – E.H.Y., M.O.B., M.K., İ.K.Y., A.E., A.Ç; Fundings – E.H.Y., M.O.B., M.K., İ.K.Y., A.E., A.Ç; Materials – E.H.Y., M.O.B., M.K., İ.K.Y., A.E., A.Ç; Data collection &/or processing – E.H.Y., M.O.B., M.K., İ.K.Y., A.E., A.Ç; Analysis &/or interpretation – E.H.Y., M.O.B., M.K., İ.K.Y., A.E., A.Ç; Literature search – E.H.Y., M.O.B., M.K., İ.K.Y., A.E., A.Ç; Writing – E.H.Y., M.O.B., M.K., İ.K.Y., A.E., A.Ç; Critical review – E.H.Y., M.O.B., M.K., İ.K.Y., A.E., A.Ç.

References

  • 1. CelebiAYalcinYErdemAZeybekCAkdenizCPolatTBSuccessful transcatheter balloon dilatation of coarctation of aorta and coil occlusion of patent ductus arteriosus in a single catheterization procedureTurk J Pediatr200749947[PubMed][Google Scholar]
  • 2. GeggelRLHijaziZMRhodesJInterventional cardiac catheterization therapy for combined coarctation of the aorta and patent ductus arteriosus:successful outcome in two infantsCathet Cardiovasc Diagn1996386770[PubMed][Google Scholar]
  • 3. HakimFHawellehAAGoussousYHijaziZMSimultaneous stent implantation for coarctation of the aorta and closure of patent ductus arteriosus using the Amplatzer duct occluderCathet Cardiovasc Interv199947368[Google Scholar]
  • 4. SadiqMMalickNHQureshiSASimultaneous treatment of native coarctation of the aorta combined with patent ductus arteriosus using a covered stentCatheter Cardiovasc Interv20035938790[PubMed][Google Scholar]
  • 5. KulkarniSVimalaJParmarRSingle therapeutic catheterization for treatment of native coarctation of aorta and large patent ductus arteriosus using a covered stentIndian Heart J2005577136[PubMed][Google Scholar]
  • 6. KrichenkoABensonLNBurrowsPMoesCAFMcLaughlinPFreedomRMAngiographic classification of the isolated, persistently patent ductus arteriosus and implications for percutaneous catheter occlusionAm J Cardiol19896787780[Google Scholar]
  • 7. RaoPSKoscikRValidation of risk factors in predicting recoarctation after initially successful balloon angioplasty for native aortic coarctationAm Heart J199513011621[PubMed][Google Scholar]
  • 8. SinghiAKSivakumarKDifferent transcatheter strategies for aortic coarctation associated with patent ductus arteriosusIndian Heart J2012644236[PubMed][Google Scholar]
  • 9. ChangZPJiangSLXuZYZhangGJHuangLJZhaoSHUse of covered Cheatham-Platinum stent as the primary modality in the treatment for native coarctation of the aortaChin Med J (Engl)201212510059[PubMed][Google Scholar]
  • 10. KhanMSMooreJWTreatment of abdominal aortic pseudoaneurysm with covered stents in a pediatric patientCatheter Cardiovasc Interv2000504458[PubMed][Google Scholar]
  • 11. VanagtWYCoolsBBoshoffDEFrerichSHeyingRTroostEUse of covered Cheatham-Platinum stents in congenital heart diseaseInt J Cardiol20141751027[PubMed][Google Scholar]
  • 12. BaruteauAEHascoetSBaruteauJBoudjemlineYLambertVAngelCYTranscatheter closure of patent ductus arteriosus:past, present and futureArch Cardiovasc Dis201410712232[PubMed][Google Scholar]
  • 13. BhalgatPSPintoRDalviBVTranscatheter closure of large patent ductus arteriosus with severe pulmonary arterial hypertension:Short and intermediate term resultsAnn Pediatr Cardiol2012513540[PubMed][Google Scholar]
  • 14. CanniffeCOuPWalshKBonnetDCelermajerDHypertension after repair of aortic coarctation--a systematic reviewInt J Cardiol2013167245661[PubMed][Google Scholar]
Collaboration tool especially designed for Life Science professionals.Drag-and-drop any entity to your messages.